Document Overview
Key Findings The P400 balloon pump (B7) completed thermal rise testing across 3 samples under continuous operation with both large and small nozzles, with all measurements recorded for reference only. Report No.: 2025041701.0 Test Date: 2025-04-17 Samples: 3 PCS (1#–3#) Ambient Temperature: 33.5 °C Test…
- Document type
- Test Report
- Prepared by
- Jessica Lin
- Published
- Last reviewed
- Topics
- Environmental Testing
Key Findings
The P400 balloon pump (B7) completed thermal rise testing across 3 samples under continuous operation with both large and small nozzles, with all measurements recorded for reference only.
- Report No.: 2025041701.0
- Test Date: 2025-04-17
- Samples: 3 PCS (1#–3#)
- Ambient Temperature: 33.5 °C
- Test Conditions: Room Temp 32 °C, Humidity 60%, Power Source 120 V
- Peak Motor Temperature: 80.8 °C (Sample B7-3#)
- Peak Temperature Rise: 47.3 K (Sample B7-3# motor)
- Verdict: For Reference Only (not a pass/fail determination)
Test Overview
| Field | Details |
|---|---|
| Test Type | Environmental Testing — Reliability Test (温升 / Thermal Rise) |
| Test Standard | Internal Control Standard (内控标准) |
| Report No. | 2025041701.0 |
| Test Date | 2025-04-17 |
| Product No. / Name | B7 — Balloon Inflation Pump (P400) |
| Manufacturer | ETENWOLF |
| Sample Quantity / No. | 3 PCS — 1#, 2#, 3# |
| Test Purpose | Evaluate the operating temperature of the samples |
| Test Conditions | Room Temp: 32 °C, Humidity: 60%, Power Source: 120 V |
| Ambient Temp (Recorded) | 33.5 °C |
Test Equipment & Setup
| Parameter | Value |
|---|---|
| Power Source Voltage | 120 V |
| Ambient Temperature (Set) | 32 °C |
| Relative Humidity | 60% |
| Ambient Temperature (Recorded during test) | 33.5 °C |
| Nozzle Configurations Tested | Large nozzle (大气嘴) and Small nozzle (小气嘴) |
| Measurement Points | Motor (马达), Internal air duct (内部风道), Large nozzle (大气嘴), Small nozzle (小气嘴) |
| Test Reference Standard | / (Internal control standard — 内控标准) |
Test Procedure
- Pre-Test: Confirm ambient temperature and humidity conditions (32 °C, 60% RH). Record baseline ambient temperature. Prepare 3 samples (B7-1#, B7-2#, B7-3#) and connect to 120 V power source.
- Procedure: Operate each unit continuously with both large and small nozzles attached. Monitor and record temperatures at four measurement points — motor, internal air duct, large nozzle, and small nozzle — until readings stabilize (temperature rise plateau reached).
- Post-Test: Record the maximum temperature (°C) and temperature rise (K) at each measurement point for each sample. Ambient temperature at time of test is noted per sample.
- Acceptance Criteria: No formal pass/fail threshold specified (internal reference standard). Results are recorded for reference only (供参考).
Test Data & Results
| Sample No. | Measurement | Motor Temp (马达温度) | Internal Air Duct Temp (内部风道温度) | Large Nozzle (大气嘴) | Small Nozzle (小气嘴) | Ambient Temp (环境温度) °C |
|---|---|---|---|---|---|---|
| B7-1# | Max Temperature (°C) | 75.2 | 74.3 | 69.1 | 71.2 | 33.5 |
| Temperature Rise (K) | 41.7 | 40.8 | 35.6 | 37.7 | — | |
| B7-2# | Max Temperature (°C) | 80.1 | 76.5 | 75.6 | 72.4 | 33.5 |
| Temperature Rise (K) | 46.6 | 43.0 | 42.1 | 38.9 | — | |
| B7-3# | Max Temperature (°C) | 80.8 | 76.8 | 72.9 | 74.3 | 33.5 |
| Temperature Rise (K) | 47.3 | 43.3 | 39.4 | 40.8 | — |
Note: Detailed time-series data is provided in the attached data sheet (测试数据见附表). Temperature rise (K) values are calculated relative to the recorded ambient temperature of 33.5 °C.
Conclusion
| Field | Details |
|---|---|
| Result | For Reference Only (■供参考 / For reference only) |
| Standard Applied | Internal Control Standard (内控标准) |
| Test Conditions | 32 °C room temp, 60% RH, 120 V power source; recorded ambient 33.5 °C |
| Samples Tested | 3 PCS — B7-1#, B7-2#, B7-3# |
| Verdict | No formal pass/fail threshold defined. Peak motor temperature 80.8 °C and maximum temperature rise 47.3 K observed on B7-3#. All data recorded for engineering reference. |
Document Information
| Role | Name (Pinyin) | Date |
|---|---|---|
| Tested By (测试/日期) | Li Guanghua | 2025-04-17 |
| Checked By (审核/日期) | Zhang Ronggui | 2025-04-17 |
| Approved By (审核/日期) | Mai Shaoqiang | 2025-04-17 |
| Source File | B7-汽球充气泵-温升报告-2025.4.17.xls | |
| Sheet | 温升报告 (Sheet 1 of 2) | |
| Report No. | 2025041701.0 | |
Frequently Asked Questions
What does temperature rise (K) mean in this test?
Temperature rise, expressed in Kelvin (K), is the difference between the maximum measured temperature at a given point and the ambient temperature recorded during the test (33.5 °C in this case). It isolates the heat generated by the product itself, independent of the starting ambient condition, making it a reliable metric for comparing thermal performance across different environments.
Why are the results marked “For Reference Only” rather than Pass or Fail?
The test was conducted against an internal control standard with no defined numerical pass/fail threshold for temperature rise. This is common in early-stage engineering validation or design characterization testing, where the goal is to map thermal behavior rather than certify compliance against an external standard such as IEC or UL.
Why were both large and small nozzles tested simultaneously?
Different nozzle sizes affect airflow restriction and therefore motor load. Running both configurations ensures the thermal profile captures the worst-case operating scenario. Results at the large and small nozzle measurement points reflect how nozzle geometry influences heat distribution along the airflow path.
How do the three samples compare?
B7-1# ran coolest overall, with a peak motor temperature of 75.2 °C and motor temperature rise of 41.7 K. B7-2# and B7-3# were close, peaking at 80.1 °C and 80.8 °C respectively. Sample-to-sample variation of approximately 5.6 °C at the motor is within a normal range for this class of product and likely reflects minor manufacturing tolerances in motor winding and housing fit.